Background:
Every year, 1.5 million people suffer a new infection by HCV, and nearly 58 million people worldwide live with a Hepatitis C chronic infection [World Health Organisation (WHO), June 2022]. To overcome the underdiagnosis of HCV among hard-to-reach population and accomplishing WHO’s target of Hepatitis C virus (HCV) elimination by 2030, it is essential to develop new rapid and easy-to-use molecular diagnostic systems with similar performance values to the gold standard qRT-PCR. Loop-Mediated Isothermal Amplification (LAMP) arises as a promising tool for point-of-care (POC) molecular detection of viral diseases. In this work, we have validated a new diagnostic tool based on fluorescent reverse transcriptase (RT) LAMP technique, detecting HCV RNA directly from capillary blood samples in less than 50 minutes.
Methods:
125 samples from HCV infected patients [116 serum samples (genotypes 1 to 4) and 9 fresh blood samples], 27 individuals who tested negative for HCV but positive for HIV, and 11 healthy controls were analysed. Serum was collected from 50 µL of blood samples by 5 min centrifugation, incubated in lysis buffer for 10 min, and then mixed with RT-LAMP reagents. RT-LAMP reactions were run in parallel with two different primer sets: HCG124, a set of 9 primers designed to detect HCV genotypes 1, 2, or 4; and HCDN-G3, a set of 6 specific primers for genotype 3. Reaction mix was first incubated at 45ºC for 10 minutes for reverse transcription, and shifted to 68ºC for 40 minutes for LAMP reaction. Sybr-green fluorescence signal was recorded in real time.
Results:
HCG124 and HCDN-G3 primer sets detected HCV RNA in 109/116 HCV positive serum samples and in 8/9 HCV positive blood samples in less than 40 minutes. This new RT-LAMP system has a sensitivity level of 94% and 100% specificity. The limit of detection of the present system is 550 – 1000 IU/mL, lower than WHO recommended limit for HCV POC testing (3000 IU/mL). Real time measurements could be substituted by a single end-point measurement at 40 minutes using a portable fluorescent reader at POC.
Conclusions:
RT-LAMP system has showed to be highly sensitive and specific for molecular detection of HCV from less than 50 µL blood sample, easily collected by capillary punction and without a prior RNA purification step. Within less than 50 minutes, an active HCV infection can be detected at POC, implementing screening strategies among vulnerable populations as a crucial step towards HCV elimination goal.