Background:
Tenofovir diphosphate (TFV-DP) is used in most HIV prevention and treatment regimens. TFV-DP levels in whole blood correlate with efficacy, and providing regular drug level feedback (DLF) leads to higher medication adherence and improves clinical care. However, the gold-standard for DLF, liquid chromatography tandem mass spectrometry (LC-MS/MS), is centralized to a few highly equipped laboratories, leading to long delays, high costs, and limited utility in routine care. We previously developed the REverSe TRanscriptase Chain Termination (RESTRICT) assay to rapidly determine TFV-DP levels based on the drug’s inhibition of DNA synthesis by HIV reverse transcriptase (RT). Here, we demonstrate the potential of a low-cost (<$300 USD) portable fluorometer, the Harmony reader, to provide semi-quantitative RESTRICT readout in near-patient settings.
Methods:
We validated semi-quantitative fluorescence measurements on the Harmony reader with RESTRICT assays containing DNA template, primers, nucleotides, recombinant HIV RT and TFV-DP. Reactions were incubated for 30 minutes at 37 °C, and PicoGreen intercalating dye was added to quench reactions and provide fluorescence readout. Harmony fluorescence measurements were compared to those obtained on a traditional benchtop plate reader (~$30,000 USD).
Results:
The Harmony reader (Fig 1A) integrates low-cost LEDs and sensors into a 3D-printed housing, is lightweight and portable, and can be USB-powered. Harmony was able to detect TFV-DP in buffer at clinically relevant concentrations (Fig 1B). The resultant enzyme inhibition curve has a 50% inhibition concentration (IC50) of 367 nM (95% confidence interval: 317-427 nM) corresponding to 750-1000 fmol/punch, demonstrating that the RESTRICT+Harmony platform can identify patients with TFV-DP concentrations that indicate moderate adherence (4 doses/week) . Harmony measurements were highly correlated with plate reader measurements (R2 = 0.96, p<0.0001, N=120).
Conclusions:
Translation of RESTRICT fluorescent measurements to a portable reader eliminates reliance on expensive laboratory equipment and could enable deployment in near-patient settings (i.e. PrEP delivery settings, pharmacies, and clinics). Overall, RESTRICT+Harmony is rapid (<30 mins) and low-cost, and with the integration of sample preparation and a larger clinical validation, could accelerate global HIV prevention and treatment efforts by providing real-time DLF to support medication monitoring and adherence.
![]()