Background: An accurate, reliable, and affordable CD4 T cells count is essential in several parts of the world. Flow cytometry is the “gold standard” for CD4 T cell count, but this technique is expensive and requires sophisticated equipment and trained personnel. In addition, the lack of ready access to technical support and quality assurance programs limits the use of flow cytometry techniques in resource-constrained countries. Instruments are now available that can solve these problems. We have tested the new portable flow cytometer for CD4+ T cell count percentage, named CyFlow MiniPOC (Partec), and analysed its sensitivity, carry-over contamination and repeatability. Its accuracy has been compared analysing the same blood samples with 2 other systems based upon CyFlow Counter (Partec) or Cytomic FC 500 (Beckman Coulter). Methodology: Venous blood from 59 adult HIV-1+ patients (age: 25-58 years; sex: 43 males; CD4 count: 34-1,115 cells/ul; CD4%: 3.1-48.0) was collected in EDTA blood, stained with the Partec MiniPOC CD4% count kit – dry kit, and analysed within a maximum of 2 hours. CD4 T cell count and percentage were determined by the CyFlow MiniPOC instrument, equipped with a 30 mW, 532 nm laser and three optical parameters for detection of side scatter (SSC), orange and red fluorescence. CD4 T cell count and percentages were measured in parallel by CyFlow Counter and by a dual platform system based upon Cytomic FC 500 (Cytostat tetrachrome kit for mAbs) and Coulter HMX (for absolute cell count). All measures were performed in triplicate. Results: The accuracy of CyFlow MiniPOC against Cytomic FC 500 showed a correlation coefficient of 0.98 and 0.97 for CD4 T cell count and percentage, respectively (linear regression analysis). The accuracy of CyFlow MiniPOC against CyFlow Counter showed a correlation coefficient of 0.99 and 0.99 for CD4 T cell count and percentage, respectively. CyFlow MiniPoc showed an excellent repeatability: CD4 absolute number and percentage were analysed on two instruments, with an intra-assay precision below +/- 10% deviation. The sensitivity was linear in the range 0-5,000 CD4 T cells/ul. There was no effect of carry-over contamination for samples at all CD4 values, regardless of their position in the sequence of analysis. Conclusions: The cost-effective and portable instrument MiniPOC produces reliable and accurate results that are fully comparable with highly expensive dual platform systems. Indeed, data perfectly correlate with those obtained with Cytomic FC 500/Coulter HMX. Finally, using CyFlow MiniPoc permits to perform in a fast and easy way a very high number of CD4 T cells counts per day.