To meet the UNAIDS 90:90:90 targets, patients in developing countries require better access to HIV viral load (VL) testing to confirm anti-retroviral (ART) treatment success. Current laboratory-based assays are confined to cities with high infrastructure, limiting access to patients in remote settings. The SAMBA II HIV-1 Semi-Q Whole Blood (WB) Test is a portable, robust, heat-stable, point-of-care (POC) VL assay that automatically leuco-depletes 3 logs of cellular RNA and DNA, thus providing clinically meaningful results (VL <or> 1000 cp/ml) using a finger-prick blood sample.
Analytical sensitivity was determined by probit analysis on dilution series of HIV-1 standard and 11 seroconversion panels. Specificity was determined by testing 500 HIV-1 negative WB samples and those containing hepatitis B, C and HTLV-I or II as well as potentially interfering substances such as bilirubin and ART drugs. Clinical sensitivity and specificity was determined using venous and finger-prick capillary blood collected from consecutive HIV-positive patients attending clinics for routine testing in four countries (Cameroon, UK, Ukraine, Zimbabwe). 1,490 venous and 816 finger-prick samples were tested. SAMBA HIV VL results using WB were compared with to Abbott Real-time PCR assay using plasma and analysed as a binary results (<1000 cp/ml or >1000 cp/ml ±0.3 Log10). Samples with discrepant results were tested with a second gold standard HIV VL assay (Roche, Hologic or BioMerieux).
Limit of detection of the SAMBA II HIV-1 Semi-Q WB Test using Probit analysis is 1,037cp/ml (95% CI: 868-1,362cp/ml) with a 95% probability of detection at 1,000 cp/ml. Specificity was 100% in 500 HIV-1 negative samples (95% CI 99.4-100%). No cross-reaction or interference was observed with common ART drugs or samples containing other infections and potentially interfering substances. Concordance between SAMBA II WB POC assay and centralised gold standard assays in clinical samples at VL > or < 1,000 cp/ml ±0.3 Log10 is given in the Table.
The SAMBA II WB VL test is highly accurate using the WHO recommended > or < 1,000 cps/ml. Results can be obtained on site as soon as 95 minutes. Phlebotomy and centrifugation are not required, making SAMBA a simple ‘sample-in/results-out' true POC assay for remote settings