Abstract Body

Viral load monitoring of HIV-1 is critical for effective HIV-infection management. This has led to the development of different HIV-1 quantitative assays. The Aptima HIV-1 Quant Dx is a quantitative assay based on real-time transcription mediated amplification and detection, run on the fully automated and random access Panther System instrument (Hologic). The objective of this study was to evaluate correlation, linearity, reproducibility and performance characteristics of the Aptima HIV-1 Quant Dx Assay (Aptima) in comparison to the reference method, Roche cobas 6800 System (c6800).

For clinical reactivity correlation, 120 retrospective and 300 prospective collected plasma samples were tested in parallel in Aptima in side-by-side testing to the reference method cobas® HIV-1 6800. Linearity and reproducibility were assessed by replicate testing of serial dilutions of well-characterized clinical samples (5 replicates of seven dilution levels spanning the assay dynamic range), and an external proficiency panel (AcroMetrix™ HIV-1 panel). Subtype reactivity was addressed in Aptima by testing an external panel (SeraCare HIV-1 Genotype). A workflow study was conducted to define the performance characteristics: protocols were designed to test different laboratory workflows when running HIV, HCV and HBV viral load assays on 3 different molecular diagnostics instruments (c6800, Veris and Panther).

Correlation between Aptima and c6800 assays in a set of 100 clinical samples within the linear range of both techniques was 0.965. The Bland-Altman mean difference estimate between the two methods was 0.24 Log cp/mL. Both systems demonstrated linearity over the dilution range (0.997 for c6800 and 0.998 for Aptima). The intra-assay variability was 1.24% for c6800 and 1.76% for Aptima, and the mean coefficient of variation inter-assay was 1.52%. Specificity for Aptima on negative samples was 100%. Aptima allowed detection and quantification of all HIV-1 subtypes tested (A, AG, B, C, D, AE, F, G and H). Panther system allowed rapid and fully automated testing: average time to first result was 161 minutes and turn around time was the shortest at different scenarios.

The Aptima HIV-1 Quant Dx is a sensitive and reproducibility assay for monitoring of HIV-1 viral load. Random access, large capacity and improved time to result make the Aptima HIV-1 Quant Dx Assay run on the Panther System a good candidate for measuring HIV-1 viral load in HIV-1 infected patients.