Abstract Body

Blood Cryptococcal Antigen (CrAg) titer is associated with risk of cryptococcal meningitis. We describe the diagnostic performance of a novel all-in-one CrAg semi-quantitative (SQ) lateral flow assay (LFA) compared to the FDA-approved Immy CrAg LFA in the rapid diagnosis of cryptococcosis. The CrAg SQ LFA provides a semi quantitative test result in a single test in contrast to traditional semi-quantitative testing which requires serial dilutions and multiple tests.

From February to September 2019, we compared the diagnostic performance of the CrAg SQ assay (Immy) with the CrAg LFA (Immy) on 100 CSF samples, 61 serum samples collected from HIV+ persons with meningitis and 50 serum samples from HIV+ persons with CD4<100 screened for cryptococcal antigenemia. The CrAg SQ levels (1+ to 4+) were compared with CrAg LFA titers to determine the corresponding cut off CrAg SQ titer. All specimens were prospectively tested. 

Among meningitis patients, 69 had cryptococcal meningitis (68 CrAg+ by LFA and 1 false negative due to prozone with 1:1310720 CrAg LFA titer), and 31 had no cryptococcal meningitis by CrAg, culture, and India ink. The CrAg SQ on CSF had 100% (69/69) sensitivity and 100% specificity (31/31) when prospectively run on fresh specimens.  CSF CrAg titers detected ranged from 1:2 to 1:209715520. In serum, sensitivity was 100% (46/46) and specificity 100% (15/15) among meningitis patients.

Using serum in asymptomatic patients, CrAg SQ had 100% sensitivity (13/13) and 97.4% (37/38) specificity with one CrAg SQ being trace 1+ positive.  Among all specimen types, the overall CrAg SQ sensitivity was 100% (128/128) and specificity was 99% (83/84); (P=0.99 by McNemar).

Overall, CrAg SQ levels from 1+ to 5+ corresponded to increasing CrAg LFA titers (Figure). CSF CrAg SQ results of 3+ or higher were always CSF culture positive.

 

The CrAg SQ LFA appears to be an excellent semi-quantitative CrAg assay maintaining both sensitivity and specificity, with rapid stratification of patient risk. Training laboratory personnel on the usage and interpretation of the CrAg SQ is necessary.