Background:
HIV displays exceptionally high virus diversity that can impact detection by diagnostic assays which fundamentally rely on sequence conservation. Therefore, it is important to confirm their performance on the breadth of known HIV strains. We tested the m-PIMA™ HIV-1/2 Detect point-of-care (POC) assay against a diverse HIV panel consisting of group M and N, group O, Circulating Recombinant Forms (CRF), Unique Recombinant Forms (URF), and HIV-2. In silico inclusivity analysis of HIV-1 and HIV-2 sequences from NCBI was performed to predict m-PIMA detection to an even broader range of circulating strains.
Methods:
m-PIMA™ HIV-1/2 Detect (Abbott Rapid Diagnostics Jena GmbH, Germany) was performed according to the manufacturer instructions. Serum from HIV patients or virus cultures with known viral load (VL) and classified sequences were tested. The panel (n=274) consisted of HIV-1 subtypes A (n=11), B (n=19), C (n=25), D (n=13), F (n=9), F1 (n=2), G (n=5), H (n=5), J (n=3), K Basal (n=2), L (n=1), group N (n=2), group O (n=16), CRF01 (n=6), CRF02 (n=13), CRF06 (n=7), CRF09 (n=4), CRF11 (n=8), CRF13 (n=4), CRF14 (n=1), CRF22 (n=1), CRF36 (n=2), CRF37 (n=2), CRF43 (n=2), HIV-1 URFs (n=106), and HIV-2 (n=5). For [i]in silico[/i] inclusivity, 1,107,963 HIV-1 and 7,289 HIV-2 NCBI sequences were used. After removing low quality sequences or those lacking assay target regions, 53,503 HIV-1 and 68 HIV-2 sequences remained to assess assay target region percent identity.
Results:
Across the cohort m-PIMA™ HIV-1/2 Detect test detected 264/274 (96.4%) samples. Mean (min-max) VL of the panel was 3.93 (2.18-6.14) log copies/mL. Among samples with HIV VL >1000 copies/mL (VL > limit of detection of the m-PIMA™ HIV-1/2 Detect) 219/222 (98.6%) were detected. At least one panel member from each subtype/CRF and all URFs were detected. In silico analysis found only 2/53,503 (0.0037%) HIV-1 (both group O) and 1/68 (1.47%) HIV-2 (subtype F) accessions with mutations in target regions which would reduce % identity below 90%, considered the threshold to ensure detection.
Conclusions:
POC testing is a critical tool to identify HIV infections around the world. Here, we demonstrate that the m-PIMA™ HIV-1/2 Detect POC assay detects each of the major circulating HIV strains as well as rare divergent strains. In silico analysis predicted that m-PIMA™ HIV-1/2 Detect would detect the majority of HIV-1 and HIV-2 strains. These data indicate that this assay can detect the full range of HIV viral diversity.