Abstract Body

The current 4th-generation HIV diagnostic screening algorithm utilizes HIV-1/2 antigen/antibody combination immunoassays. Reactive samples are then tested by orthogonal immunoassays to confirm and discriminate between HIV-1 and HIV-2 antibodies. Additional HIV nucleic acid testing (NAT) is often required to resolve indeterminate or undifferentiated HIV seroreactivity. Undetectable HIV-2 plasma RNA (<10 copies/mL) is reported in one-third of treatment-naïve HIV-2-infected patients; as such, a second-line NAT that detects HIV-2 DNA/RNA in peripheral blood mononuclear cells (PBMC) is essential to confirm HIV-2 infection in this circumstance.

We developed and validated an HIV-2 total nucleic acid (TNA) test using the Abbott m2000 platform. Amplification of human DNA and internal control RNA was used to assess for sample quality and assay inhibitors. Assay sensitivity was evaluated by testing known copies of HIV-2 plasmid DNA mixed with 1 million PBMC. Matched plasma and PBMC samples were collected from 25 HIV-1, 30 HIV-2, 8 HIV-1/2 dual-seropositive and 25 HIV-seronegative participants. Diagnostic performance was evaluated by comparing the outcome of the HIV-2 TNA assay with the results obtained by the Abbott HIV Ag/Ab Combo assay (Combo) and the Bio-Rad Multispot HIV-1/HIV-2 Rapid Test.

The assay sensitivity was 26 TNA copies/million cells (95% CI, 18-55 copies/million cells). Thirty of thirty (100%; 95%CI lower bound 88.4%) PBMC samples from HIV-2 seropositive participants tested positive for HIV-2 TNA, including 21 samples from participants with undetectable HIV-2 plasma RNA. All thirty matching plasma samples from the HIV-2 TNA-positive participants were reactive for Combo. Among these thirty Combo-reactive samples, twenty nine were confirmed HIV-2 Ab positive and one was nonreactive by Multispot. Plasma samples from 50 HIV-2 TNA-negative individuals were confirmed non-reactive for HIV-2 Ab. Thus, the overall agreement between the HIV-2 TNA assay and the combined results of the immunoassays was 98.8% (79 of 80). The TNA assay also detected HIV-2 in 7/8 PBMC samples from HIV-1/-2 dually infected participants with undifferentiated Multispot and Geenius HIV-1/HIV-2 assays seroreactivity.

Our HIV-2 TNA assay detected HIV-2 in PBMC from serologically HIV-2 reactive, HIV indeterminate, and HIV undifferentiated individuals with undetectable plasma HIV-2 RNA, and is suitable for confirming HIV-2 infection in the CDC HIV testing algorithm.