Abstract Body

Background: The 2013 WHO guidelines for antiretroviral treatment (ART) of HIV infections recommend viral load (VL) for routine monitoring and diagnosis of treatment failure (TF) defined as > 1,000 copies/ml. Due to operational simplicity, dried blood spots (DBS) are recommended as an alternative specimen type however at a higher TF diagnostic threshold of 3-5,000 copies/ml. DBS based VL testing has been extensively evaluated in comparison with plasma based testing however performance has varied. We assessed DBS against plasma based VL performance under field laboratory conditions using the Cobas Ampliprep/TaqMan Version 2.0 (CAP/CTM), a common platform for screening for suspected virological TF. Methodology: EDTA whole blood (WB) specimens were obtained at three health facilities from 823 ART patients after ≥6 months on ART when blood was submitted for routine or targeted plasma VL tests at the Namibia national reference laboratory. DBS were prepared from these blood specimens within 6 hours after blood draws. DBS were dried overnight and packaged according to the WHO recommended method. DBS VL was determined using CAP/ CTM HIV-1 version 2.0 test kits following the manufacturer’s instructions. The assay has a lower detection limit of 2.60 log10 copies/ml. DBS VL data were analyzed using plasma VL as reference standard. Results: Using DBS (n=823) the clinical sensitivity, specificity, positive (PPV) and negative predictive value (NPV) with 95% confidence intervals (CI) were 0.99 (0.96 – 0.99), 0.55 (0.52 – 0.59), 0.33 (0.29 – 0.38) and 0.99 (0.99 – 1.0), respectively, at a threshold of 5,000 copies/ml; and 0.99 (0.97 – 1.0), 0.26 (0.22 – 0.29), 0.29 (0.26 – 0.33) and 0.99 (0.90-1.0), respectively, at a threshold of 1,000 copies/ml. The prevalence (95% CI) of TF was 0.18 (0.16 – 0.21) and 0.23 (0.21 – 0.26) in plasma and 0.55 (0.51 – 0.58) and 0.80 (0.77 – 0.83) in DBS at the 5,000 and 1,000 copies/ml thresholds, respectively. Conclusions: Low specificity and PPV of the CAP/CTM on DBS at both evaluated threshold levels indicate high rates of false positive results for virological TF may occur if this approach were used for routine ART patient monitoring and TF diagnosis. Even use of the more conservative DBS specific TF threshold of 3-5,000 copies/ml recommended by the 2013 WHO guidelines may result in uneccessary ART regimen changes for patients responding favorably to current treatment. These results add to the published literature highlighting suboptimal specificity and PPV of DBS based VL testing. As such, recommendations for use of DBS for routine VL monitoring may require further evaluation. Additional research is needed to better understand the effectiveness of these approaches in routine programmatic settings, including the performance of other assay versions and types.