Background: Quantification of cell-associated HIV-1 RNA is critical to understanding how HIV-1 latency is established and maintained in a reservoir of infected cells during antiretroviral therapy. Droplet digital polymerase chain reaction (ddPCR) can detect and precisely quantify the target nucleic acids without the need for a calibration curve. Previous studies demonstrated that ddPCR is generally comparable to traditional qPCR methods for the quantification of both total HIV-1 DNA and episomal 2-LTR circles. In this study, we developed a reverse transcriptase (RT) ddPCR assay and compared to quantitative reverse transcriptase-PCR(qRT-PCR) for detection and quantification of unspliced cell-associated HIV-1 RNA in both HIV-uninfected cell lines spiked with HIV-infected cells and clinical samples from patients on antiretroviral therapy (ART) with very low-level viremia. Methodology: To evaluate the ability of RT-ddPCR to measure cell-associated HIV-1 RNA, serially diluted (2.5 to 10^8 copies/10 ul RNA input) in vitro transcribed HIV-1 RNA was quantified by RT-ddPCR and qRT-PCR. We analyzed 8 replicates of the 100-copy standard to assess the accuracy and reproducibility of RT-ddPCR. Levels of cell-associated HIV-1 RNA were determined in clinical samples from patients on suppressive ART, and from MT2 cells spiked with the latently-infected cell line ACH2. Results: RT-ddPCR accurately quantified HIV-1 RNA transcripts; the dynamic range was 10 to 10^5 copies per 10 ul RNA input (R2 > 0.99), which represented similar sensitivity as qRT-PCR. The 100-copy input showed that the percentage of positive droplets most closely correlated with the expected copy number. For samples with input RNA of 10^3 to 10^5 copies, the copies detected was ~10% lower than expected; for samples with <100 copies of input RNA the calculated copy number was slightly higher than expected (mean 1.9-fold difference). Linearity decreased significantly when greater than 10^5 copies due to saturation of positive droplets formation. HIV-1 RNA measured by RT-ddPCR and qRT-PCR in samples from co-culture of serially diluted reactivated ACH2 cells together with MT2 cells lines were highly correlated (R2=0.9668). HIV-1 RNA was detected by RT-ddPCR in 14/20 samples (range, 4-284 copies/million PBMC) and by qRT-PCR in 13/20 samples (range, 6-479 copies/million PBMC), respectively. Conclusions: The RT-ddPCR assay detected and accurately quantified cell-associated HIV-1 RNA with a dynamic range down to 10 copies per reaction. RT-ddPCR demonstrated similar accuracy and sensitivity compared to qRT-PCR in the measurement of cell-associated HIV-1 RNA in both cell-lines spiked with HIV-infected cells and in patient samples. RT-ddPCR is a promising tool for the quantification of HIV-1 cell-associated RNA.