A non-sputum triage test to rule out tuberculosis disease has been identified as a high-priority need for diagnostic development to reach the End-TB targets of the World Health Organization (WHO). The target product profile defines a minimum of 90% sensitivity and 70% specificity for the test to be used by first-contact providers to identify patients who need further confirmatory testing. A combinatory score based on a novel 3-gene host-signature has shown promise in discriminating TB disease from other respiratory illnesses and healthy controls. Cepheid (Sunnyvale, CA, USA) has developed an early prototype GeneXpert PCR test (‘Xpert Prototype’), that quantifies relative mRNA-levels of the 3-gene signature in a patient whole blood sample.
Whole blood from symptomatic people living with HIV (PLHIV) in South Africa were collected from February 2016 to August 2017 and biobanked in PAXgene tubes. The accuracy of the Xpert Prototype on these biobanked samples was compared against a comprehensive microbiological reference standard (culture and Xpert® MTB/RIF). The performance was also compared against Xpert® MTB/RIF alone, as Xpert will be the most likely confirmatory assay used in programmatic settings in high-burden countries. We depict results in ROC curves and for pre-set cut-points based on performance targets set for a triage test by the World Health Organization.
Of the 201 patients included, 67 were culture-positive. At a cut-point chosen to maximize the Youden index, sensitivity was 77.6% (95% Confidence Interval [CI] 66.3–85.9), specificity was 92.2% (CI 86.3–95.7) and the AUC was 0.89 (CI 0.83-0.94) against the comprehensive reference standard. Considering the Xpert-Prototype as a triage test (fixed sensitivity value closest to 90%), the corresponding specificity was 55.8% (CI 47.2-64.1). Comparing to Xpert® MTB/RIF alone as a confirmatory test at fixed value of sensitivity closest to 90% (90.6%), the Xpert Prototype specificity was 85.9% (CI 79.3-90.7). Considering the Xpert Prototype as a stand-alone diagnostic test, at a specificity of 95%, the test achieves a sensitivity of 65.7% (CI 53.7-75.9).
In this first accuracy study of a novel blood-based host-marker assay on a commercial platform, we show the possible value of the assay for triage and potentially also for diagnosis, when a sputum sample is difficult to obtain, as often the case in PLHIV.