Abstract Body

Background: Recent revisions of HIV testing algorithms have aimed to minimise the time to confirmation of the diagnosis without losing sensitivity and specificity which has clear advantages to both patients and clinicians. We hypothesised that the positive predictive value (PPV) of the Abbott Architect HIV Ag/Ab Combo Assay increases with the signal to cut-off (S/CO) ratio and that a S/CO ratio could be identified with a sufficiently high PPV to give a positive result to patients without waiting for the results of supplementary testing.

Methods: All testing episodes were extracted from the laboratory database between March 2006 and March 2014. Samples came from a wide range of adult clinical services including medical and surgical inpatient departments, other laboratories, sexual health clinics, infectious diseases clinics, a women’s hospital, IVF clinics, general practitioners and prison health services. Positive tests were classified as true positive or non-true positive depending on synchronous and subsequent supplementary testing with a Western Blot (WB), a HIV p24 Ag assay (confirmed by neutralization), a HIV Ab assay (tested with an alternate EIA) and/or HIV viral load. The data were randomly allocated to two equally sized samples – a train sample and a test sample.

Results: Of 138,911 testing episodes, 3,705 had a positive result. The true positive group included samples with : 1) a synchronous positive WB (N=1,989), 2) an initial negative or indeterminate WB followed by a positive WB within 6 months (N=79), 3) a negative or indeterminate WB and at least two positive results of the p24 Ag assay, the HIV Ab assay or the HIV viral load (N=435), 4) a negative or indeterminate WB and a positive p24 Ag assay without a HIV viral load result available (N=38).The highest non-true S/CO ratio in the train sample was 151.17 and when this was applied to the test sample it had a PPV of 100% and a sensitivity of 67.4%. S/CO ratios of 100 and 50 had a PPV and sensitivity of 99.7% and 73.8% and 99.0% and 80.6% respectively. The frequency distribution of the S/CO ratios of true and non-true positives in the total sample is shown in figure 1.

Conclusions: We have shown that knowledge of the S/CO ratio allows for the test result of the Abbott Combo to be used immediately in clinical decision making. Given a sufficiently high S/CO ratio, the clinician will be able to deliver a test result to the patient without waiting for further confirmatory testing with a high degree of clinical certainty.

viewer?TAG_ACTION=SECURE_FILE_DOWNLOAD&V