Cryptococcal antigen (CrAg) titers are an important prognostic indicator in HIV-positive patients with cryptococcal infection and could potentially be used to stratify treatment. Current titration methods are expensive and labor intensive. A novel semi-quantitative (SQ) CrAg test has been developed that can provide an indication of CrAg titer using a simple dipstick lateral flow assay (LFA). We performed a study to evaluate the performance of the SQ-CrAg assay against the standard CrAg LFA in patients with HIV-associated cryptococcal infection in Gaborone.
Residual EDTA blood specimens from sequential HIV-positive patients undergoing routine CD4 testing with CD4 counts of ≤200 cells/ml were screened through the reflex CrAg-screening program in Botswana using both the IMMY CrAg LFA and the novel SQ-CrAg LFA. The sensitivity and specificity of the SQ-CrAg in the reflex CrAg screening cohort were determined relative to the standard CrAg LFA. To further validate the SQ assay known CrAg+ EDTA blood samples from a prior CrAg-screening study and a CM treatment trial were also tested with both assays. Serial dilutions were performed for all CrAg+ samples and re-tested with the standard LFA to determine titres. SQ titers and conventional titers were compared. All testing was performed by two independent blinded investigators and inter-rater reliability assessed using the Kappa coefficient.
692 sequential samples were screened using both assays; 43 (6.2%) were IMMY CrAg LFA-positive. Using this standard CrAg LFA as a reference, the overall sensitivity and specificity of the novel SQ-CrAg LFA were 93.0% (95%CI 80.9 – 98.5%) and 93.8% (95%CI 91.7-95.6%) respectively. A further 180 known CrAg+ samples were tested and the combined results used to evaluate the SQ-CrAg quantification. Median (IQR) CrAg titers for SQ-CrAg 1+, 2+, 3+, and 4+ bands were 1:10 (1:5 – 1:20), 1:40 (1:20 – 1:80), 1:640 (1:160 – 1:2560), and 1:5120 (1:2560 – 1:20480) respectively (Figure 1). Inter-rater agreement in titer assessment was excellent at 98.2%, with a kappa coefficient of 0.96, p<0.001.
Overall sensitivity and specificity of the novel IMMY SQ-CrAg assay were high in a cohort of HIV-positive individuals with CD4 counts ≤200 cells/ml undergoing reflex CrAG screening. An SQ titre of 3+ or greater corresponded to a titer of >1:160 which has previously been shown to be associated with increased mortality.