Abstract Body

In most resource limited countries with a high HIV burden, current Viral Load (VL) testing technologies are based at the centralized laboratory levels. New HIV VL point-of-care (POC) tests hold the potential to simplify ART monitoring and bring diagnostics closer to the patient. The objective of the study was to perform a laboratory validation of the diagnostic accuracy of the GeneXpert® HIV-1 Quant assay compared to the predicate laboratory reference standard BioMérieux NucliSens Version 2.0 Easy Q/Easy Mag (NucliSens v2.0) for VL testing

EDTA anticoagulated plasma samples from 392 consenting patients on ART were tested for HIV-1 VL at the National Microbiology Reference Laboratory (NMRL) in Harare, Zimbabwe from June-September 2015. Paired samples were tested with each reference and index test following the manufacturer’s instructions and laboratory SOPs. The technician performing the Xpert VL test was blinded to the result of the reference test. Statistical analysis was conducted in STATA 13.0 and included sensitivity, specificity, positive predictive value (PPV), linear regression and Bland-Altman (BA) analysis.

Of the 392 samples were tested in the Xpert® HIV-1 Quant assay, 375 produced a VL result (3.4% true error rate). The sensitivity of the Xpert® HIV-1 Quant assay in detecting HIV VL as at 1000copies/ml threshold compared to the NucliSens v2.0 assay was 98.2% [95% CI: 93.8-99.8], specificity 97.7% [95%CI: 95.1-99.2] and PPV 94.8% while NPV 99.7%. Bland-Altman mean bias between the two methods was -0.22 log10 copies/ml [95% CI: -0.292 to -1.52] and the LOA for the bias were -1.11 to 0.66; p-value=0.125. Spearman’s correlation coefficient, R2=0.952. Concordance between the two assays at 1000copies/ml threshold was 97.8%. Only 1.6% of patients (6 of 375) and 0.5% (2 of 375) were misclassified above and below this threshold respectively by the index assay compared to predicate assay. The system proved easy-to-use with minimum sample preparation and handling time, with an average daily capacity of 16 samples per 8hour day

Xpert® HIV-1 Quant gives HIV VL results comparable to NucliSens v2.0 assay.  Findings from this study support the consideration of Xpert® HIV-1 Quant testing for the roll-out of HIV VL monitoring and may contribute towards reaching the ambitious UNAIDS target of 90% ART patients with undetectable VL