Background: Pre-treatment loss to follow-up can be a barrier to effective antiretroviral treatment (ART) programs. CD4 point-of-care (POC) testing may reduce this by improving linkage to care. However, an implementation challenge is accuracy when operated by lay health workers (HIV testing counselors and community health workers) in field settings. We validated the accuracy of Pima’s CD4 POC device (PCD4D) when performed by lay health workers (LHW) versus laboratory technicians (LT), with LT-performed FACSCalibur as the gold standard, using capillary and venous blood from HIV-1 infected patients. Methodology: In Phase I, we compared PCD4D performed in the laboratory by LT against FACSCalibur for 280 venous specimens. In Phase II we compared PCD4D performed by LHW versus LT from 147 paired capillary and venous specimens, and compared these to LT-performed FACSCalibur (venous and capillary). Statistical analyses included agreement (Bland-Altman tests), correlation concordance coefficient (CCC), and sensitivity plus specificity for classification as above/below FACSCalibur-defined CD4 count of 350 cells/mm3 (Kenya’s current threshold for ART initiation). Results: PhaseI: PCD4D compared favorably with FACSCalibur on venous specimens; venous sensitivity and specificity at 350 cells/mm3 were both 88.9% (Table 1). Phase II: Good performance was observed for venous PCD4D results from both LHW (CCC=87, bias-86cells/μl) and LT CCC=91,bias- 66cells/μl). Capillary PCD4D also compared favorably: LHW (CCC=93,bias-44cells/μl), LT (CCC=94,bias-36cells/μl). Venous sensitivity and specificity were 100 and 98% respectively for LT, and 94 and 95% for LHW. Using capillary blood, the sensitivity and specificity were 79 and 98% (LT) and 86 and 99% (LHW). Conclusions: POC CD4 results were comparable to FACSCalibur for both venous and capillary specimens, when operated by both lay health workers and lab technicians. POC CD4 testing has potential to enhance HIV care and treatment programs without burdening laboratory technicians in resource limited settings.